Press the Enter key to confirm the weight value and begin the span adjustment.
: Locate a small screwdriver or a pen to access the internal hardware calibration switch (SW1). Step-by-Step Avery Berkel L126 Calibration Process
When the display reads CAL 0 or flashes 0 , press the Enter key.
Navigate through the menu using the arrow keys or the key until the display reads CAL 0 or ZERO . avery berkel l126 calibration
Power on the indicator for at least 15 to 30 minutes. This allows the internal load cell electronics and strain gauges to reach thermal equilibrium.
: The display will then read "load 10" (or another specific weight value). Gently place your standard calibration weight on the center of the platform and press the Test button .
Power the unit back on. Locate and press the hardware micro-switch on the left side of the primary circuit board. The screen will display a flashing S . Press the Enter key to confirm the weight
For complex technical issues or to register for warranty support, visit the Avery Berkel Service Portal .
| | Meaning / Display | | :--- | :--- | | Overload | The weight on the scale exceeds its capacity. The display will typically show a series of dashes: – – – – – – – – . | | Underload | The load cell signal is too low, often indicating a disconnected or faulty load cell. The display will also typically show – – – – – – – – . | | Out of Zero Range | The scale is beyond its automatic zero-tracking range (±2% by default). The display may show \- - - - - - - - . |
: Use the red Zero button to scroll through units until your preferred measure (e.g., "ounce" or "kg") appears. Confirm by pressing the blue Test button . Navigate through the menu using the arrow keys
The following steps are standard for Avery Berkel/Avery Weigh-Tronix L-series indicators:
The Avery Berkel L126 is a staple in the weighing industry, known for its reliability in industrial and retail environments. However, like any precision instrument, it requires periodic calibration to maintain accuracy and compliance with weights and measures regulations.
If the scale continues to display errors or fails to hold its calibration, it may indicate a damaged load cell, failing mainboard components, or structural binding underneath the platter. In these instances, contact an authorized Avery Berkel service technician for advanced diagnostic support.
: Once the display reads "done" , remove the weight. Press the Test button to restart the scale and return to normal weighing mode.
Return the physical calibration switch to the or LOCK position. The scale will reboot into normal weighing mode.